Fluke
Fluke 726 Precision Multifunction Calibrator
Model No :
Fluke 726
Part Code :
2452080

Overview

Fluke 726 device is specially designed for the process industry.This is an easy-to-use field calibrator that supports broad workload coverage, calibration power, and unsurpassed accuracy needed by process professionals. The 726 interprets results without the use of calculator and stores measurement data for later analysis.

Features
- Designed specifically for the process industry
- Delivers precise measurement and calibration source performance with accuracies of 0.01%.
- Stores up to eight calibration results in memory for later analysis
- Offers HART mode that inserts a 250 ohm resistor in mA measure and source for compatibility with HART instrumentation

Major Applications
- Basic measuring, testing and regulation techniques for setting and verifying test installations and instrumentation
Technical Specifications
- Voltage DC :- 30.000 V; 0.01% + 2 mV
- Current DC :- 24.000 mA; 0.01%+ 2 mA
- Resistance :- 0.0 to 4000.0 O
- Current DC :- 24mA; 0.02% + 2 counts
- Frequency :- 15kHz; 0.05% + 0.01 kHz
- Pressure :-Accuracy from 0.025% of range using any of 29 pressure modules
- Temperature: RTD types :- 8 types ; Resolution :- 0.01°C (0.01°F)
- Temperature: TC types :- 12 types ; resolution :- J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
- Source Accuracy:
• Voltage DC: 20.000 V 0.01% + 2 mV
• Current DC:24.000 mA (Source); 0.01% + 2 µA
• Resistance: Range, accuracy (percent of reading + floor) and excitation current 5 to 4000 O, 0.015% + 0.1 O
• Frequency: 10.0 to 15.0 kHz ; 0.5%
• Waveform: 5 V p-p squarewave, -0.1 V offset



General Specifications
- Operating temperature:-10°C to +55°C
- Storage temperature: -20°C to +71°C
- Operating altitude: 3000 m
- Safety Specifications Agency approvals: EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992